Heating Unit 2000

Heating Unit 2000

The Heat­ing Unit 2000 sup­plies the large Pecon incu­ba­tors with heat­ed air up to max. 30°C above ambi­ent temperature.

Description:

  • The Heat­ing Unit 2000 sup­plies the large PeCon incu­ba­tors with heat­ed air up to max. 30°C above ambi­ent temperature
  • The Heat­ing Unit 2000 auto­mat­i­cal­ly adapts to the impressed line voltage
  • High oper­a­tional safety:
    If the tem­per­a­ture of the inter­nal heat­ing block exceeds 170°C, the heater is per­ma­nent­ly deac­ti­vat­ed (ther­mal pro­tec­tion Switch); after an excess tem­per­a­ture shut-off, the user can restart the device
  • The speed of the fan is per­ma­nent­ly monitored:
    If it falls below a thresh­old, the heater is switched off
  • New con­stant power system:The released heat­ing power is dynam­i­cal­ly adapt­ed to the cur­rent line volt­age; as a result, there is no over­heat­ing, a more sta­ble incu­ba­tor tem­per­a­ture and a high­er heat­ing power even with low line voltages
  • For the con­trol of air tem­per­a­ture, the Tem­p­Con­troller 2000-1 or 2000-2 is required

 

Performance:

Typ­i­cal fast heat­ing of a large PeCon incu­ba­tor by the Heat­ing Unit 2000:

  1. Heat­ing up to 37.0°C with­in 30-45 min.
  2. Lev­el­ling with very low set­point devi­a­tion between 45 and 60min.
  3. Sta­ble oper­a­tion after 60-75min.


Long time sta­ble heat­ing of a large PeCon incu­ba­tor over sev­er­al days. Max­i­mum air tem­per­a­ture devi­a­tion: +/- 0.1°C.

Specifications:

AC operating voltage AC 100...115/230 V - autorange; 50 - 60 Hz
Power consumption 615 VA max.
Operating voltage (from TempController 2000-2) DC 24 V
Power consuption (from TempController 2000-2) 250 mA max.
Mains fuse 250 V 5.0 AT(H)
Max. heating power typ. 22-25°C above ambient temperature; dependent on the incubator type and the tube length
Thermal protection switch Switching temperature: > 170°C Resetting temperature: < 80°C
Dimensions (WxHxD) 190 x 154 x 215 mm
Weight approx. 3.7 kg
Operating temperature + 5°C up to + 40°C
Rel. humidity max. 80 %, non-condensing
Altitude max. 2000 m above sea level
Overvoltage category 2
Pollution degree 2
Protection class I
Heating Unit 2000

Compatibility:

  • all devices of the PeCon Series 2000 (and the Zeiss Series S)

Application Examples:



Data protection overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.

Please note our Privacy statement.