CO2-O2-Controller 2000

CO2-O2-Controller 2000

CO2-O2 con­trol device to achieve a sta­ble, con­trolled increase of the CO2 and a decrease of the O2 con­cen­tra­tion with­in PeCon incu­ba­tors and CO2-Covers. The CO2-O2-Con­troller 2000 can be used with all devices of the Incu­ba­tion Sys­tem 2000 or Incu­ba­tion Sys­tem S.

Description:

  • CO2-O2 con­trol device for the gen­er­a­tion of a defined CO2  and O2 con­cen­tra­tion and a sub­se­quent pressure-less sup­ply of incu­ba­tors and CO2-Covers
  • CO2 com­ing from a CO2 gas cylin­der is added to the ambi­ent air
  • A built-in CO2 sen­sor per­ma­nent­ly mon­i­tors the CO2 con­cen­tra­tion in the air stream and reports the value to the con­trol unit which oper­ates a pro­por­tion­al valve and reg­u­lates the amount of CO2 added to the air flow
  • Nitro­gen com­ing from a N2 gas cylin­der is added to the ambi­ent air in order to decrease the amount of oxy­gen (O2)
  • A built-in long last­ing zir­co­ni­um based O2 sen­sor per­ma­nent­ly mea­sures the O2 con­cen­tra­tion of the efflu­ent gas mix and reports the value to the con­trol unit which oper­ates a pro­por­tion­al valve and reg­u­lates the amount of nitro­gen which is added to the air flow
  • A pump con­trolled by the con­trol unit pumps in ambi­ent air into the com­bined meter­ing cham­ber of the gas mea­sur­ing sen­sors in addi­tion to the con­trolled gas inflow of CO2 and N2. No need for pres­sur­ized air supply!
  • A Humid­i­fi­ca­tion Bot­tle (500 ml) is includ­ed in all PeCon oxy­gen reduc­ing con­trollers, com­pen­sat­ing the strong dry­ing effect of the added gas
  • Rec­om­mend­ed range for the CO2 set­point value from 0.0 Vol-% up to 20.0 Vol-%: Res­o­lu­tion of the CO2 dis­play: 0.1 Vol-% Res­o­lu­tion of the inter­nal CO2 loop con­trol: 0.01 Vol-%
  • Rec­om­mend­ed range for the O2 set­point value from 0.0 Vol-% up to 21.0 Vol-%: Res­o­lu­tion of the O2 dis­play: 0.1 Vol-% Res­o­lu­tion of the inter­nal O2 loop con­trol: 0.01 Vol-%
  • The com­mu­ni­ca­tion with a win­dows PC is car­ried out via the USB con­nec­tion which per­mits an easy updat­ing of the firmware

Specifications:

Operating voltage 90 - 260 V AC, 50...60 Hz
Power consumption max. 50 W
Mains fuse 250 V 4AT(H)
CO2 supply 1-2 bar (15-30 psi); max. 2.5 bar (36 psi) technical CO2, purity 99 %
CO2 display range 0.0 up to 100.0 Vol-% resolution 0.1 Vol-%
CO2 loop control range (setpoint) 0.0 up to 20.0 Vol-%
CO2 loop control accuracy typ. ± 2 % of setpoint value (e.g.: 5.0 ± 0.1 Vol-%) internal resolution 0.01 Vol-%
Heat-up Phase of CO2 sensor approx. 15-30 min
N2 supply 1-2 bar (~15-30 psi); max. 2.5 bar (~36 psi) technical N2, purity >99%
O2 display range
O2 loop control (setpoint)
O2 loop control accuracy
Heat-up phase of O2 sensor
Outlet for gas mixture
Dimensions (WxHxD) 190 x 130 x 255 mm
Weight approx. 5.25 kg
Operating temperature + 5°C up to + 40°C
Rel. humidity max. 80 %, non-condensing
Altitude max. 2000 m above sea level
Overvoltage category 2
Pollution degree 2
Protection class I
CO2-O2-Controller 2000

Requirements:

  • none

Compatibility:



Data protection overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.