TempController 2000-2

TempController 2000-2

The tem­per­a­ture con­trol unit with two chan­nels elec­tri­cal­ly heats var­i­ous PeCon com­po­nents. The Tem­p­Con­troller 2000-2 can be used with all devices of the Incu­ba­tion Sys­tem 2000 or Incu­ba­tion Sys­tem S.

Description:

  • Tem­per­a­ture con­trol unit with two equal chan­nels to elec­tri­cal­ly heat var­i­ous com­po­nents of the manufacturer
  • For the use togeth­er with other com­po­nents of the Incu­ba­tion Sys­tem 2000 (PeCon) or the Incu­ba­tion Sys­tem S (Zeiss)
  • The addi­tion­al port allows e.g. the oper­a­tion of a LED Illu­mi­na­tion inside the incubator
  • Both chan­nels are set up with the name of the con­nect­ed com­po­nent on the oper­a­tor interface
  • Can be oper­at­ed with the but­tons at the front side as well as exter­nal­ly via USB
  • The con­trol char­ac­ter­is­tic of the dig­i­tal PID loop con­trol can be adapt­ed to the con­nect­ed com­po­nent: For this pur­pose, 8 para­me­ter sets which can be freely assigned to one of the two heat­ing chan­nels are available
  • Res­o­lu­tion of tem­per­a­ture dis­play: 0.1°C.; res­o­lu­tion of inter­nal loop con­trol: 0.01°C
  • Range for set­point val­ues from 0.0 up to 60.0°C; this range can be lim­it­ed by spe­cial para­me­ter sets to pro­tect spe­cif­ic com­po­nents (e.g. Objec­tive Heater 2000)
  • A power sup­ply pack with a wide input volt­age range allows the oper­a­tion of the Tem­p­Con­troller 2000-2 with all com­mon AC line volt­ages with­out man­u­al adaptation
  • Self-resetting over­load pro­tec­tion in the chan­nels 1 and 2

Specifications:

AC operating voltage AC 90-260 V, 50... 60 Hz
Power consumption 200 W max.
Mains fuse 250 V T3.15 A(H)
Output DC 24 V; max. 4 A per channel; max. 4.2 A combined power
Temperature display 0.0 up to 99.9°C resolution 0.1°C
Loop control range (Setpoint) 0.0 up to 60.0°C
Loop control accuracy typ. ± 0.15% of setpoint value (e.g.: 37.0 ± 0.05°C) internal resolution 0.01°C
Dimensions (WxHxD) 190 x 130 x 255 mm
Weight approx. 4.25 kg
Operating temperature + 5°C up to + 40°C
Rel. humidity max. 80 %, non-condensing
Altitude max. 2000 m above sea level
Overvoltage category 2
Pollution degree 2
Protection class I
 
TempController 2000-2

Requirements:


  • none

Compatibility:



Data protection overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.